| Temperature Range | |
| Korrox - III Salt Spray Chamber | Adjustable from ambient to +50°C/ +122°F |
| Korrox - III Salt Spray Chamber | Korrox - III – 450 Litres |
| Standard Salt Spray Chamber | Korrox Corrosion III- 450 Litres | |
| Premium CASS cum Salt Spray | Korrox Corrosion III- 450 Litres | |
| Chamber Capacity | 450 Litres/ 8.83cu ft. | |
| Mounting Format | Floor Standing | |
| Chamber external dims, max | W | 2200 |
| D | 1150 | |
| H | 1550 | |
| Chamber internal dims, max | W | 1000 |
| D | 700 | |
| H | 750 | |
| Chamber Construction | Robust Construction with Fibre reinforced body (Triple wall) | |
| Electricity supply | ||
| Korrox - III Salt Spray Chamber | Single Phase | |
| Korrox - III Salt Spray Chamber | 15 (VAC) Voltage and 50 (HZ) Frequency | |
| Water Requirements | 15 (VAC) Voltage and 50 (HZ) Frequency DM/deionized for supply to air saturator and making salt solution. Air saturator requires a continuous water connection 1.5-5.0 bar (21-73 psi). | |
| Air Requirements | Clean dry & dust free compressed air, 4.0 to 6.0 bar (58-87psi) along with 240 Ltrs (8.5cu ft.) per minute flow. | |
| Fume Exhaust Requirements | 2 inch exhaust pipe is connected which should be drained outside Test area or Lab premises. | |
| Drain Requirements | 1 inch drain pipe is connected which should be drained outside building. | |
| Operating Environmental Conditions | +18°C to 23°C (+64° to 73°F), 85% RH (ambient conditions) max. |