Full list of standards
Cyclic Corrosion Chamber Specifications for Presto range of Equipment.
| Wetting Mode | Temperature Range | Adjustable from + 5°C above ambient to + 70°C / +158°F |
| Humidity Range | Controlled at 95% - 100%RH | |
| Salt Spray Mode | Temperature Range | Adjustable from + 5°C above ambient to + 60°C / +140°F |
| Salt Spray Fog Collection Rates | Adjustable from 0.5 to 2.5 ml per 80 cm² per hour | |
| Drying Mode | Temperature Range | AAdjustable from + 5°C above ambient to + 70°C / +158°F |
| Humidity Range | Uncontrolled | |
| Controlled Humidity Mode | Temperature / Humidity Range | Adjustable from + 5°C above ambient to + 70°C / +158°F |
| Cyclic Corrosion Test Chambers | ![]() |
CCT - 450 |
| Chamber Capacity | 15.8 cu. Ft. | |
| Mounting Format | Floor Standing |
| Chamber External Dimensions, max | W | 2200 |
| D | 1100 | |
| H | 1150 | |
| Chamber Internal Dimensions, max | W | 1000 |
| D | 700 | |
| H | 750 | |
| Salt Solution Reservoir Capacity | H | 60 Litres |
| Chamber Construction | H | Fibre Reinforced Plastics (FRP) |
| Electricity supply | Single Phase Supply | |
| 15 (VAC) Voltage and 50 (HZ) Frequency | ||
| Water | DM/deionized for supply to air saturator and making salt solution. Air saturator requires a continuous water connection 1.5-5.0 bar (21-73 psi) | |
| Air | Clean dry & dust free compressed air, 4.0 to 6.0 bar (58-87psi) along with 240 Litres (8.5cu ft.) per minute flow. | |
| Exhaust | 2 inch exhaust pipe is connected which should be drained outside Test area or Lab premises. | |
| Drain | 1 inch drain pipe is connected which should be drained outside building. | |
| Operating Environmental Conditions | +18°C to 23°C (+64° to 73°F), 85% RH (ambient conditions) max. |