Full list of standards
Cyclic Corrosion Chamber Specifications for Presto range of Equipment.
Wetting Mode | Temperature Range | Adjustable from + 5°C above ambient to + 70°C / +158°F |
Humidity Range | Controlled at 95% - 100%RH | |
Salt Spray Mode | Temperature Range | Adjustable from + 5°C above ambient to + 60°C / +140°F |
Salt Spray Fog Collection Rates | Adjustable from 0.5 to 2.5 ml per 80 cm² per hour | |
Drying Mode | Temperature Range | AAdjustable from + 5°C above ambient to + 70°C / +158°F |
Humidity Range | Uncontrolled | |
Controlled Humidity Mode | Temperature / Humidity Range | Adjustable from + 5°C above ambient to + 70°C / +158°F |
Cyclic Corrosion Test Chambers | CCT - 450 | |
Chamber Capacity | 15.8 cu. Ft. | |
Mounting Format | Floor Standing |
Chamber External Dimensions, max | W | 2200 |
D | 1100 | |
H | 1150 | |
Chamber Internal Dimensions, max | W | 1000 |
D | 700 | |
H | 750 | |
Salt Solution Reservoir Capacity | H | 60 Litres |
Chamber Construction | H | Fibre Reinforced Plastics (FRP) |
Electricity supply | Single Phase Supply | |
15 (VAC) Voltage and 50 (HZ) Frequency | ||
Water | DM/deionized for supply to air saturator and making salt solution. Air saturator requires a continuous water connection 1.5-5.0 bar (21-73 psi) | |
Air | Clean dry & dust free compressed air, 4.0 to 6.0 bar (58-87psi) along with 240 Litres (8.5cu ft.) per minute flow. | |
Exhaust | 2 inch exhaust pipe is connected which should be drained outside Test area or Lab premises. | |
Drain | 1 inch drain pipe is connected which should be drained outside building. | |
Operating Environmental Conditions | +18°C to 23°C (+64° to 73°F), 85% RH (ambient conditions) max. |