• STANDARD YOU FOLLOW

  • Full list of standards

Cyclic Corrosion Chamber Specifications

Cyclic Corrosion Chamber Specifications for Presto range of Equipment.

Wetting Mode Temperature Range Adjustable from + 5°C above ambient to + 70°C / +158°F
Humidity Range Controlled at 95% - 100%RH
Salt Spray Mode Temperature Range Adjustable from + 5°C above ambient to + 60°C / +140°F
Salt Spray Fog Collection Rates Adjustable from 0.5 to 2.5 ml per 80 cm² per hour
Drying Mode Temperature Range AAdjustable from + 5°C above ambient to + 70°C / +158°F
Humidity Range Uncontrolled
Controlled Humidity Mode Temperature / Humidity Range Adjustable from + 5°C above ambient to + 70°C / +158°F
Cyclic Corrosion Test Chambers CCT - 1000
Chamber Capacity 35.3 cu. Ft.
Mounting Format Floor Standing
Chamber External Dimensions, max W 2750
D 1650
H 1600
Chamber Internal Dimensions, max W 1600
D 800
H 550
Salt Solution Reservoir Capacity H 60 Litres
Chamber Construction H Fibre Reinforced Plastics (FRP)
Electricity supply Single Phase Supply
15 (VAC) Voltage and 50 (HZ) Frequency
Water DM/deionized for supply to air saturator and making salt solution. Air saturator requires a continuous water connection 1.5-5.0 bar (21-73 psi)
Air Clean dry & dust free compressed air, 4.0 to 6.0 bar (58-87psi) along with 240 Litres (8.5cu ft.) per minute flow.
Exhaust 2 inch exhaust pipe is connected which should be drained outside Test area or Lab premises.
Drain 1 inch drain pipe is connected which should be drained outside building.
Operating Environmental Conditions +18°C to 23°C (+64° to 73°F), 85% RH (ambient conditions) max.